If the light incubator suddenly does not work, check the fuse tube

Constant temperature and humidity incubator.jpg

Shanghai Xinzhuang Instrument Co., Ltd. Xinzhuang (LNB) brand intelligent light incubator is widely used in microbial tissue cell culture, seed germination, seedling experiment, plant culture, etc., can simulate different environmental climate conditions.

Common faults and maintenance of intelligent light incubators:

1. The incubator housing should be reliably grounded.

2. The incubator should be placed in a cool, dry, well-ventilated place away from heat and sunlight. Place it smoothly to prevent noise from occurring.

3. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a 50mm gap on the side of the box and at least 300mm on the top of the box.

4. When the incubator is transported, repaired and maintained, it should avoid collision, shaking and shaking; the maximum inclination is less than 45 degrees.

5. The instrument suddenly does not work. Please check whether the fuse tube (after the box) is burnt out and check the power supply.

6. When the incubator is working in refrigeration, it is not advisable to make the difference between the temperature inside the box and the ambient temperature greater than 25 degrees.

Common faults in intelligent lighting incubators

1. The compressor is not cooled, the temperature inside the incubator is not uniform, or the circulating air outlet is frozen.

2. In the balanced mode, the compressor is always in working condition, the cooling effect is better and the compressor can be effectively maintained.

3. If the temperature is reduced to a certain value when working in a single system, the temperature will fluctuate greatly when the door is opened or closed. Causing the compressor to be ordinary

4, the start is easy to shorten its service life.

5, the temperature is not uniform:

6. It may be that the discharged items are too dense or the circulating fan inside the box does not work.

7. If you use long light, the light bulb will easily heat up and the temperature will be uneven.

The outlet is frozen:

Each refrigerated instrument has a defrost cycle during which the compressor does not cool. The ice on the condenser is melted by the heating element. The melted water is likely to collect at the air outlet and eventually cause ice to form.

Shanghai Xinzhuang Instrument Co., Ltd. produces a series of light incubators, high-quality, high-performance product quality for the national colleges and universities, scientific research institutions. Perfect service allows users to buy with confidence, use comfortably. Welcome new and old customers to inquire !

Http://

Presure seal Gate Valve has two type of sealing way, flexible wedge disc and parallel slab disc which normally used in the high pressure main streaming pipeline or high pressure bypass water feeding pipeline etc. The main function is  to cut off or cut on the medium controlled by electric actuator or handwheel. The sealing principle of pressrue seal gate valve is sealing surface tight off under pressure and they are widely applicated in petrochmical, thermal power plant. 

Pressure Seal Gate Valve

Pressure Seal Gate Valve,Flange Gate Valve,Power Station Valve,Wedge Disc Gate Valve

Yongjia South Trading Co.,Ltd , https://www.n-lvalve.com

This entry was posted in on