Canine distemper virus IgG antibody (CDV-IgG) ELISA detection kit

Canine distemper virus IgG antibody (CDV-IgG) ELISA detection kit

user's Guide

Detection principle

The kit uses double antibody one-step sandwich enzyme-linked immunosorbent assay (ELISA). To the coated microwells pre-coated with canine distemper virus IgG antibody (CDV-IgG) capture antibody, add specimens, standards, and HRP-labeled detection antibodies in sequence, incubate and wash thoroughly. The color is developed with the substrate TMB, which is converted into blue under the catalysis of peroxidase and into the final yellow under the action of acid. The color depth is positively correlated with the canine distemper virus IgG antibody (CDV-IgG) in the sample. Measure the absorbance (OD value) at 450nm with a microplate reader to calculate the sample concentration.

Sample collection, processing and storage methods

1. Serum: Use a test tube that does not contain pyrogens and endotoxins. Avoid any cell stimulation during the operation. After collecting blood, centrifuge at 3000 rpm for 10 minutes to quickly and carefully separate serum and red blood cells.

2. Plasma: EDTA, citrate or heparin anticoagulation. Take the supernatant by centrifugation at 3000 rpm for 30 minutes.

3. Cell supernatant: centrifuge at 3000 rpm for 10 minutes to remove particles and polymers.

4. Tissue homogenate: Add tissue to the right amount of saline and mash. Take the supernatant by centrifugation at 3000 rpm for 10 minutes.

5. Preservation: If the sample is not tested in time after collection, please aliquot it in one dose and freeze it at -20 ℃ to avoid repeated freezing and thawing. Thaw at room temperature and ensure that the sample is thawed evenly and fully.

Bring your own items

1. Microplate reader (450nm)

2. High-precision sampler and tip: 0.5-10uL, 2-20uL, 20-200uL, 200-1000uL

3. 37 ℃ thermostat

Operation notes

1. Store the kit at 2-8 ° C and equilibrate at room temperature for 20 minutes before use. The concentrated washing liquid taken out of the refrigerator will have crystals, which is a normal phenomenon. The water bath is heated to completely dissolve the crystals before use.

2. The slats not used in the experiment should be immediately returned to the ziplock bag, sealed (dry at low temperature) and stored.

3. The standard dilution can be regarded as a negative control or blank; the sample after pretreatment does not need to be diluted, just take 10μL and add it.

4. Perform the incubation operation strictly in accordance with the time, amount of liquid and sequence indicated in the manual.

5. Shake all liquid components thoroughly before use.

Kit composition

name

96-well configuration

48 hole configuration

Remarks

Microplate

12 holes × 8

12 holes × 4 strips

no

Standard product (1200IU / L)

0.6mL

0.6mL

Dilute according to the instructions

Standard dilution

6mL

3mL

no

Sample diluent

6mL

3mL

no

Detection antibody-HRP

10mL

5mL

no

20 × washing buffer

25mL

15mL

Dilute according to the instructions

Substrate A

6mL

3mL

no

Substrate B

6mL

3mL

no

Stop solution

6mL

3mL

no

Sealing film

2 sheets

2 sheets

no

Instructions

1 serving

1 serving

no

Ziplock bag

1

1

no

Note: The standard product is diluted sequentially with standard product diluent: 1200, 600, 300, 150, 75, 37.5IU / L.

Reagent preparation

Dilution of 20 × washing buffer: 1:20 dilution of distilled water, that is, 1 part of 20 × washing buffer plus 19 parts of distilled water.

Washing method

1. Manually wash the plate: throw away all the liquid in the hole, fill each hole with the washing liquid, leave the liquid in the hole after standing for 1 min, pat dry on the absorbent paper, and wash the plate 5 times in this way.

2. Automatic plate washing machine: Inject 350μL of washing liquid into each well, soak for 1min, and wash the plate 5 times.

Steps

1. Take out the required slats from the aluminum foil bag after equilibrating at room temperature for 20 min. The remaining slats are sealed with a ziplock bag and put back at 4 ° C.

2. Set up standard wells and sample wells, add 50μL of standard products of different concentrations to the standard wells;

3. Add 10μL of the sample to be tested first, and then add 40μL of the sample diluent;

4. Then add 100 μL of detection antibody labeled with horseradish peroxidase (HRP) to each well of the standard and sample wells, seal the reaction well with a sealing plate, and incubate for 60 min in a 37 ° C water bath or incubator.

5. Discard the liquid, pat dry on the absorbent paper, fill each well with the washing solution, let stand for 1min, shake off the washing solution, pat dry on the absorbent paper, and repeat washing the plate 5 times (you can also wash the plate with a washing machine)

6. Add 50 μL of substrate A and B to each well, and incubate at 37 ° C in the dark for 15 minutes.

7. Add 50μL of stop solution to each well, and measure the OD value of each well at 450nm wavelength within 15min.

Result judgment

Draw standard curve: In the Excel worksheet, the standard product concentration is used as the abscissa, and the corresponding OD value is used as the ordinate. The standard product linear regression curve is drawn, and the concentration value of each sample is calculated according to the curve equation.

Kit performance

1. Accuracy: The correlation coefficient R between the linear regression of the standard product and the expected concentration is greater than or equal to 0.9900.

2. Sensitivity: The minimum detection concentration is less than 1.0 IU / L.

3. Specificity: Does not cross-react with other soluble structural analogs.

4. Repeatability: The coefficients of variation within and between panels are less than 15%.

5. Storage: 2-8 ℃, protected from light and moisture.

6. Validity: 6 months

Disclaimer

1. The kit is for research use only, and should not be used for clinical or human experiments, otherwise all the consequences will be borne by the experimenter, and the company will not be responsible.

2. Strictly follow the instructions, and the experimenter violates the instructions, and the consequences will be borne by the experimenter.

Stationery Tape

Stationery tape is mainly used to fix all the little imperfections in the office, home and school. Write on it with pen, pencil or marker, every word is clear. With smooth surface, easy to cut.  With different color and different size, the thickness of stationery adhesive tape is from 45 to 65 micron, the width is from 12mm to 24mm, the length is 20 to 100 yards. Could make custom stationery tape. 

Stationery Tape

Stationery Tape,Crystal Clear Stationery Tape,Stationery Bopp Tape,Colored Stationery Tape

Dongguan Yalan Packing Materials Co., Ltd. , https://www.yalanpack.com

This entry was posted in on